Model 6510e/6530 Power Analyzers

SpecificationsAnalog OutputExternal ShuntsData SheetManual
 
Power Analyzers The Magtrol 6510e Single-Phase and 6530 Three-Phase Power Analyzers are easy-to-use instruments ideal for numerous power measurement applications. From DC to 100 kHz AC, the 6510e and 6530 measure volts, amps, watts, volt-amps, frequency, crest factor, Vpeak, Apeak and power factor in one convenient display. They may be used either as stand-alone instruments or in conjunction with any Magtrol Hysteresis, Eddy-Current or Powder Brake Dynamometer; any Magtrol Dynamometer Controller and M-TEST Software for more demanding motor test applications.
 
Features
  • Single/Three-Phase Capabilities: Equipped to handle single (6510e) or three-phase (6530) power measurements
  • Ranges: Up to 600 Vrms @ 20 A continuous duty
  • Interfaces: RS-232 & IEEE-488 Interfaces
  • Data Transfer Rates: Up to 100 per second
  • Accuracy: Up to 0.1%
  • Vacuum Fluorescent Display: High-quality, easy-to-read, customizable readout displays volts, amps, power and power factor
  • Measurement: Continuous or cycle-by-cycle
  • Bandwidth: DC up to 100 kHz
  • Input Power: Accepts 120/240 Vrms, 60/50 Hz power at 20 VA max
  • Auto Ranging: Automatically scales instrument for most accurate range
  • Isolation: 1000 Vrms to earth, 750 line-to-line
  • Average: Displays running average of amps, volts and watts
  • Peak Hold: Stores the highest value read. Values can include amps, watts and volts in any combination
  • Analog Outputs: Plug-in module that provides 12 channels of analog output corresponding to volts, amps and watts.
  • External Sensor Input
  • Calibration Certificate: NIST Traceable
  • Rack Mounting: 19" (482.6 mm) with handles
 
Applications
  • Motors and Drives
  • Lighting Fixtures and Ballasts
  • Office Equipment
  • Household Appliances
  • Power Tools
  • HVAC Equipment
  • Calibration of Test and Measuring Instruments
  • Switching Power Supplies
  • Power Inverters
  • Transformers
The 6510e's/6530's data transfer rate makes them ideal for static and dynamic tests.
 
Back to top
 

Technical Specifications
  VOLTAGE INPUT CURRENT INPUT EXTERNAL SENSOR INPUT POWER
Ranges 30 V, 150 V, 300 V, 600 V 1 A, 5 A, 10 A, 20 A 50 mV, 250 mV, 500 mV, 1.0 V ---
Crest Factor 2.7 @ full scale input 1.7 @ full scale input 2.4 @ full scale input ---
Impedance 2 megohm 12 milliohm 17 kilohm ---
Display Range 5 digits with 10 mV resolution 5 digits with 1 mA resolution 5 digits with 0.1 mV resolution 6 digits with 1 mW resolution
ACCURACY
DC ±(0.1% Reading + 0.2% Range) 0.4% of VA range
5 Hz - 500 Hz ±(0.1% Reading + 0.1% Range) 0.2% of VA range
500 Hz - 10 kHz ±((0.015 × F(kHz)% Reading) + 0.3% Range) 0.6% of VA range
10 kHz - 100 kHz 0.6% + (0.03% × F(kHz)% of VA range)
>100 kHz N/A measurement band limited to DC-100 kHz
 
FREQUENCY
Source V1,A1, V2,A2, V3,A3, LINE or EXT
(For Vx and Ax, source input must be >10% of range setting)
Accuracy 20 to 500 Hz 0.05%
EXT. Input BNC non-isolated, earth ground referenced
Input Impedance 100 kilohm
Voltage Level TTL / CMOS
Maximum Voltage 50 V
 
Back to top
 

Analog Output Option
The Analog Output is a plug-in module designed specifically for use with the 6510e Single-Phase Power Analyzer and 6530 Three-Phase Power Analyzer. It provides up to 12 channels (3 for the 6510e, 12 for the 6530) of analog output corresponding to volts, amps and watts. Each output is capable of, and calibrated to, ± 10 volts. With the analog output option, the user may apply a scale factor to all outputs. Amps scaling is applied to all amp channels, volts scaling is applied to all volt channels and watts scaling is applied to all watt channels.
 
Features
  • Simple Installation: Plug-in module automatically configures power analyzer for operation
  • Full Scale Output Range: ± 10V maximum output (user scalable)
  • Up to 12 Channels: 6510e Single-Phase Power Analyzer (3 channels), 6530 Three-Phase Power Analyzer (12 channels)
  • 14-Bit Resolution
  • Frequent Output Updates: Updates every 5 milliseconds
  • Output Short Protected
  • Factory Calibrated: Calibration data/values stored on board (EEPROM)
  • Industry Standard 25-Pin Connection
 
Applications
The Analog Output can be used along with a 6510e or 6530 Power Analyzer to output information to a strip recorder or interface with a data acquisition system. The Analog Output is also used as a safety feature to help trigger alarm events.
SPECIFICATIONS
Resolution 14 bits
Temperature
Coefficient
4 ppm FSR/°C (typical),
20 ppm FSR/°C (maximum)
Output Range ± 10 V maximum
Number of Channels 12
Basic Accuracy 3 least significant bits
Output Update Rate
(all channels)
200 samples/second (5 ms)
Ordering Information
6510e Single-Phase Power Analyzer
6510e-01 Single-Phase Power Analyzer with Analog Output Option
6530 Three-Phase Power Analyzer
6530-01 Three-Phase Power Analyzer with Analog Output Option
 
Back to top
 

External Shunts

If currents continuously reach above 20 amps, an external sensor must be used. Magtrol offers three different types of external shunts for use with Magtrol Models 6510e or 6530 High-Speed Power Analyzers.

The HA Series shunt, designed for power cable hook-up utilizing 3/8-16 screws, has an insulating base and can be mounted on conductive surfaces. The LAB Series shunt also comes with an insulating base, along with thumb screws for the power leads and knurled nuts on the sensing terminals. The FL Series is a relatively small bus, bar-mounted shunt with large end blocks. The comparatively large end blocks and short spacing aid in cooling and allow the shunt to operate in more extreme environments. All shunts are calibrated on equipment with current certifications traceable to N.I.S.T.

External Shunts
 
Specifications
  HA Series LAB Series FL Series
Rated Current 50 A to 500 A 750 A to 1000 A 2000 A to 3000 A
Output 50 mV
Bandwidth DC to 60 Hz
Accuracy ± 0.25%
(± 0.1% accuracy with calibration certificate
is available for an additional cost)
Operating Temperature For optimum accuracy, temperature of shunt
should be 30 °C to 70 °C
Available Models
Type P/N Amps
HA 004640 50
HA 004641 100
HA 004642 150
HA 004643 200
HA 004644 250
HA 004645 300
HA 004646 400
HA 004647 500
LAB 004648 750
LAB 004649 1000
FL 005214 2000
FL 005286 3000
Sensor Substitution Wiring Connection
Shunt Configurations
NOTE: For the 6530 Three-Phase Analyzer, any (or all) of the Amps inputs may be replaced with a shunt for any of the wiring modes.
 
Back to top
 